Line 25: | Line 25: | ||
* If a STS has occurred, you set this new audiogram as the ''Revised Baseline Audiogram''. | * If a STS has occurred, you set this new audiogram as the ''Revised Baseline Audiogram''. | ||
− | == | + | == Sampling == |
− | === 8111 === | + | === Substance Codes === |
+ | |||
+ | ==== 8111 ==== | ||
Hearing Conservation Program (Action Level @ 50% dose / 85 dBA) | Hearing Conservation Program (Action Level @ 50% dose / 85 dBA) | ||
− | === 8110 === | + | ==== 8110 ==== |
Engineering Controls (PEL @ 100% dose / 90 dBA) | Engineering Controls (PEL @ 100% dose / 90 dBA) | ||
+ | |||
+ | === Edge eg5 Noise Dosimeter (3M) === | ||
+ | |||
+ | ==== Detection Management Software ==== | ||
+ | ★ ''This software is required when changing settings on the device, including setting date/time. There is no way to do this from the device itself.'' | ||
+ | |||
+ | ==== Procedures for setting/checking configuration ==== | ||
+ | # Set each of the eg5 dosimeters into the charging base. | ||
+ | # Power on each dosimeter. (NOTE: Don't skip this part. Turn it on!) | ||
+ | # Ensure the USB cable is plugged into both the charging base and the computer. | ||
+ | # Open the 3M DMS software | ||
+ | There is a configuration file saved in the IH Shared folder on the H: drive: | ||
+ | |||
+ | H:\IH Shared\OSH Reference\Equipment\Configuration Files\Edge_Dosimeter_Config_ky.xml | ||
[[Category:Health]] | [[Category:Health]] |
Revision as of 17:50, 19 June 2017
- OSHA uses a 5 dB doubling rate (aka, exchange rate). For every 5dB increase, maximum time until 100% dosage halves.
- Peak: At no point during an employee's work shift should they be exposed to a sound pressure level over 140dB C-weighted.
- NOTE: I previously had non-weighted (otherwise known as Z-weighting) because no OSHA reference material or publications seen by me had reference to weighting as it applies to peaks. However, the ACGIH and others set this limit with C-weighting.
- Error Factor → The error of the dosimeter must be factored in to the consideration of a violative condition.
- For example, the dosimeters used at KYOSH are Type II dosimeters. This indicates an error of 2 dB.
- A Type I dosimeter has an error of only 1 dB.
Noise Standards
For Noise Exposure — 1910.95(a) and (b)
- Exchange (Doubling) Rate: 5 dB
- Criterion Level: 90 dBA
- Threshold: 90 dBA*
For Hearing Conservation — 1910.95(c) (General Industry amendment; not Construction)
- Exchange (Doubling) Rate: 5 dB
- Criterion Level: 90 dBA
- Threshold: 80 dBA*
*See OSHA Tech Manual Section III, Chapter 5, I.A.3. and I.A.4.
Standard Threshold Shift (STS)
An STS is a change in hearing threshold relative to the baseline audiogram of an average of 10 dB or more at 2000, 3000, and 4000 Hz in either ear.
- When a STS has occurred and the employee’s hearing level is at least 25 dB above audiometric zero in the same ear, the hearing loss is recordable on 300 logs.
- If a STS has occurred, you set this new audiogram as the Revised Baseline Audiogram.
Sampling
Substance Codes
8111
Hearing Conservation Program (Action Level @ 50% dose / 85 dBA)
8110
Engineering Controls (PEL @ 100% dose / 90 dBA)
Edge eg5 Noise Dosimeter (3M)
Detection Management Software
★ This software is required when changing settings on the device, including setting date/time. There is no way to do this from the device itself.
Procedures for setting/checking configuration
- Set each of the eg5 dosimeters into the charging base.
- Power on each dosimeter. (NOTE: Don't skip this part. Turn it on!)
- Ensure the USB cable is plugged into both the charging base and the computer.
- Open the 3M DMS software
There is a configuration file saved in the IH Shared folder on the H: drive:
H:\IH Shared\OSH Reference\Equipment\Configuration Files\Edge_Dosimeter_Config_ky.xml