Revision as of 17:06, 26 October 2017
The ToxiRAE Pro Single-Gas Detectors, manufactured by RAE Systems, were purchased by KYOSH in 2017. A replacement was needed for the aging and irreparable Quest SafeLog 100s, which were used for carbon monoxide (CO) and nitrogen dioxide (NO2) detection. The units themselves and parts were no longer being manufactured and existing part stocks were quickly dwindling. Because KYOSH responds relatively frequently to imminent danger calls concerning carbon monoxide, it is imperative to employee safety that KYOSH continuously has functional detectors. The SafeLogs could no longer provide that assurance.
Enter the ToxiRAE Pro. It is a complete replacement for the SafeLog detectors, as there are ToxiRAE Pros for CO detection and NO2 detection.
Physical Interface
TO DO: Location of lights, type of sensor installed, buttons, speaker, sensor, belt clip, charging/communication contacts
On-Screen Interface
TO DO: Items on main screen, menu, settings
Sampling
When using the ToxiRAE to detect either carbon monoxide or nitrogen dioxide, the following apply to both gases.
- Whether you're screening or personal air monitoring, use the OSHA-93 Direct Reading Report for any direct read devices, including the ToxiRAE Pro.
Configuration Settings
Calibration /
"Span" |
High
Alarm |
STEL
Alarm |
TWA
Alarm |
IDLH1 | |
---|---|---|---|---|---|
CO | 50 | 200‡ | YD | 50† | 1200 |
NO₂ | 5.0 | 5.0† | 1.0‡ | YD | 20 |
1 Not a configurable setting. For reference only.
YD There are no regulatory or suggested limits, so what you enter here is at your discretion.
† OSHA limits
‡ NIOSH recommended limits